How is Closure Managed with Impella® Heart Pumps?

Manual compression is the standard closure technique with Impella heart pumps.1 Key steps of the manual compression technique are:2

  • Sheath removal: Operator applies gentle pressure 1-2 cm above the access site and removes the sheath.
  • Hemostasis: Immediately, operator applies firm downward pressure for 15-30 minutes (5 minutes of full pressure, 5 minutes of 75% pressure, 5 minutes of 50% pressure and 5 minutes of 25% pressure) or until hemostasis is achieved.

 

 

View more resources

References

  1. Burzotta, F., et al. (2015). Int J Cardiol, 201, 684-691.
  2. Impella Certification Workbook 2018.

IMP-395

This information is intended for use by customers, patients, and healthcare professionals in [region] only. We recognize that the Internet is a global communications medium; however, laws, regulatory requirements, and product information for medical products can vary from country to country. The product information included here may not be appropriate for use outside [region], and the information from other sites you visit may not be appropriate for use in [region].